Medical Ultrasound & Electronics / Signal Processing

Waveform Index Evaluation using an Electronic Injection Phantom

 D.J. Potter, X.F. Li,
Dept. of Medical Physics & Medical Engineering, Royal Infirmary of Edinburgh.

( "IEEE Engineering in Medicine and Biology" Vol 14, No.1, January/February 1995 )

In Action 2:
Injecting a pre-recorded real waveform.

The audio waveform injected was generated by the onboard audio-frequency oscillator, allowing a highly reliable calibration signal from which to guage the Doppler phantom's displayed velocity response.
Return to Main Paper