| |
| |
A versatile Electronic Signal Injection (ESI) phantom for use in Doppler Ultrasound Quality Assurance testing has been developed within the Dept. of Medical Physics and Medical Engineering, Royal Infirmary of Edinburgh, Scotland. The phantom allows injection of real, prerecorded Doppler waveforms which facilitates testing of waveform index calculation packages. An internal, discrete audio frequency generator also allows a check to be made on the accuracy of a scanners' frequency abscissa calibration.
Sideband suppression is typically in excess of 50 dB over the range 1 - 10 MHz for audio shifts of between 300 Hz and 7kHz, allowing easy identification of directional discrimination faults (channel crosstalk) for a wide variety of scanner frequencies. The feasability of performing waveform index calculation package checking as part of a quality assurance (QA) routine with this phantom is assessed.
Medical ultrasound can generally be catagorised into two fields, these being image generation and blood flow rate estimation using the Doppler effect. Typically a small, hand-held transducer generates a scanning directional beam of ultrasound which maps out a 2-D image. At an interface between differing tissue types, variations in densities and speed of sound causes some of the beam energy to be scattered (reflected) back towards the transducer. An image is constructed by displaying the intensity of the echo received against time for each angular increment of the beam's sweep area. Further, if the beam encounters a moving scattering source, such as blood cells, the shift in frequency of the scattered beam caused by the Doppler effect can be used to determine the velocity of the scatterer from the Doppler formula
| (1.1) |
where fd is the frequency shift induced in a reflected sound beam of frequency fo when it encounters a scattering source moving at velocity v at an angle q (the `Doppler' angle) to the wavefront. The speed of sound in the medium is denoted by cs.
The differing flow rate across a blood vessel combined with scattering effects between the blood cells produce a spread of frequency shifts at any one instant, so that in `Doppler' mode (frequency shift versus time) the scanner displays an entire waveform. Combining the information from the image (blood vessel diameter, beam-blood vessel interception angle) with that from the Doppler effect one can deduce the mean blood flow rate through the vessel, a figure of significant clinical importance.
The situation is, however, not as simple as is suggested above. Two sources of error can combine to produce an erroneous blood flow rate. Firstly, several assumptions have to be made about the mean tissue sound speed, the flow profile of the blood in the vessel, etc. For non-turbulant, steady flow through a straight, cylindrical vessel far from any bends, the mean blood flow rate may be deduced simply by knowing the vessel diameter and the velocity in the centre of the vessel where the blood flow is fastest (maximum frequency of the waveform). One must also be absolutely sure that the ultrasound scanner produces reliable information. The blood vessel diameter, as often measured using electronic callipers (`imaging' mode), and the frequency shift (`Doppler' mode) are of prime importance in determining the mean flow rate through a vessel and both must be accurately known.
To this end, various test objectsd or `phantoms' have been developed to test both the physical assumptions mentioned above and the accuracy of the ultrasound scanners. Flow phantoms typically circulate blood mimicking fluid (such as a glycerine and water mixture) through a plastic tube embedded in tissue mimicking material. Fine scattering particles suspended within the fluid produce a signal similar to that from blood, and the fluid is circulated by some form of pump. Although such phantoms serve well in checking the accuracy of `imaging' mode parameters (beam-blood vessel angle and vessel diameter, etc.) and flow dynamic parameters, for QA work they suffer several drawbacks. Flow phantoms are generally bulky, extremely non-portable and take a great deal of time to prepare for use ( such as the preparation of synthetic blood). Furthermore, their performance characteristics change with time and are not therefore easily reproducable. All of this combines to make them unsuitable for checking Doppler mode parameters.
A simpler idea is embodied in the `string phantom', which uses a simple rotating thread on a system of pulley wheels to act as a line scattering source1. This phantom has the advantage of portability and ease of use, and can perform frequency abscissa checking in Doppler mode i.e. rotating the string at a constant velocity corresponding to a known Doppler frequency shift (for a set string-ultrasound beam angle q) and checking the scanner waveform to ensure the displayed frequency shift matches the theoretical one. For the testing of ultrasound scanners in Doppler mode of operation, string phantoms have one main setback - the inability to produce realistic Doppler waveforms for waveform index calculation tests. Varying the string speed can produce a variable maximum-frequency envelope, although a precise specification of the envelope is often difficult without a computer controlled motor to drive the string. Even then, as will be discussed more fully later in this paper, the advantages to be gained by doing so are questionable.
In both the flow phantom and the string phantom, one is faced with the inability to solely test the scanner electronics. For one thing, the physical dimensions of a transducer's active area introduce an additional angular dependence through the vessel-ultrasound beam angle q, a parameter of critical importance in the Doppler equation 1.1. This results in `geometrical' spectral broadening2 of an ideal single frequency into a frequency spectrum, and clearly can be a significant source of error in determining the maximum frequency of a waveform, particularly at high Doppler angles q. This effect becomes worse as the transducer is brought closer to the scattering source. Accurate determination of the Doppler angle in itself is non-trivial given that many transducers have slightly curved faces and were designed for clinical use rather than laboratory experiments. Furthermore, a good operator can only determine q with an accuracy of approximately 3o.
Thus, if one wishes to test a scanner's frequency abscissa calibration, for instance, one finds measurement errors tend to swamp any error arising from the scanner electronics alone. This problem is eliminated with a new type of phantom which is independent of angular effects, the electronic phantom.
The concept of using electronic phantoms for Doppler Ultrasound QA work was introduced by Evans in 19893. The principle of operation is to return a frequency shifted version of the ultrasonic signal received from the scanner under test using a transmit / receive transducer pair, acoustically coupled by a sculpted block of perspex. This is typically accomplished with the aid of electronic devices which perform the act of multiplication between their two inputs, and which are known as `mixers'. The whole process has been termed `Electronic Signal Injection', or ESI. The induced frequency shift of typically a few kHz is interpreted by the scanner as a Doppler shifted version of its own emitted signal. As the Doppler shift is generated electronically, no angle measurements are necessary and that principal source of uncertainty is eliminated.
Evans used a straightforward filtering technique to remove the unwanted sideband generated by the mixing process and obtain a simulated unidirectional Doppler source. If filters are to be effective they require sharp cut-off characteristics to avoid unwanted attenuation effects over the frequency range spanning the desired sideband.
More recently, Wallace4 in Newcastle has described a wideband ESI phantom capable of injecting simple waveforms such as a sawtooth frequency sweep. By using a wideband single sideband (SSB) generator they have obtained a sideband suppression ranging between 18 and 37 dB over a range of 1 - 10 MHz, for audio shifts of up to 10 kHz. Although a considerable improvement over the prototype phantom of Evans et al, the broadband nature of the Newcastle SSB generator is responsible for their rather low sideband suppression, which is disadvantageous for scanner directional descrimination assessment. There is clearly an oppertunity for improvement in the field of electronic signal injection systems.
The elimination of angular sources of error not only make electronic phantoms ideal for frequency abscissa checking, but for another highly desireable QA procedure, but one which has been neglected somewhat in the literature - the assessment of the waveform index calculation packages on Doppler scanners. As these packages expect to operate on real Doppler waveforms, the many advantages of ESI phantoms3,4 would be incomplete without the capability to inject realistic waveforms. These factors provided the motivation for the construction of an improved ESI prototype phantom during the period May 1992 to May 1993.
Figure 1.1 illustrates the principle of operation of the Edinburgh ESI
phantom. The acoustic ultrasound signal from the scanner is converted
into a radio frequency (RF) electrical signal by means of an input
transducer attached to the ESI phantom. The signal is then split into
two signals shifted in phase by [(p)/2] radians. Each RF signal
is then electronically mixed with an audio frequency signal in a device
known as a balanced `modulator' or `mixer', which performs the mathematical
operation of multiplication. The audio frequency signals, as might
be expected, are also [(p)/2] radians out of phase with each other.
The output of each modulator consists of two (radio) frequencies, termed sidebands, which are symmetrically displaced by an amount equal to the audio frequency from the scanner's frequency, termed the carrier. This technique forms the basis for AM radio transmission, from which we borrow the relevant terminology. Summation of the modulator outputs produces a single sideband (SSB) output necessary to simulate a unidirectional Doppler source.
By introducing [(p)/2] phase shift circuits at the appropriate places, the unwanted sideband is cancelled out, giving rise to the name of this technique - the `phasing method'. This technique is capable of producing very high suppression of the undesired sideband through destructive interference between two antiphase phase RF signals. One disadvantage is that this technique cannot easily be implemented over a wide bandwidth, as the phantom of Wallace et al demonstrated. Instead, we have opted to choose twelve scanner frequencies between 2 MHz and 10 MHz in use, according to a survey, in hospitals within Lothian and Fife Regions.
In the Edinburgh phantom precise quadrature phase shifts at each scanner
frequency are achieved by using parallel tuned R-L-C circuits for
the ultrasonic signals, as depicted in figure 1.2. At the resonant
frequency of each such circuit, two signals in quadrature phase with one
another but of equal amplitude are drawn from the inductor and capacitor
using high impedance buffers. The inductor is slightly variable to allow
fine tuning of each circuit for precise quadrature phase shifting.
Although successful cancellation of the undesired sideband ideally requires identical respective signal amplitudes and exact 900 phase shifts, an error bound of 1o is widely regarded as an upper limit for good sideband suppression5. Sideband suppression figures suggest we have obtained a phase error between RF signals of less than 1o, in comparison with the results of Wallace et al who quote phase errors of 4o with their broadband RF filter. By using multiple parallel tuned circuits of the type shown in figure 1.2 we have successfully devised a compromise to obtain maximum sideband cancellation at frequencies of practical use to us with a minimum of electronic complexity.
Waveform injection is achieved by means of a second, audio phasing board generously donated by Duncan Cairns, manager of Dynamic Imaging Ltd, Livingston. This board uses a combination of active and passive filters to produce two signals of equal amplitude but in quadrature phase with one another from any audio input frequency in the range 300 Hz to 7 KHz. Technically, the board consists of a `polyphase' network6,7, a network of resistors and capacitors connected in a ring configuration, the output of which is a fourfold set of quadrature signals. The reader is referred to the references for further details which are out of place here.
Further flexibility is provided by an onboard audio oscillator
microchip which produces quadrature phase audio
signals. These may be fed directly to the mixers, allowing single frequency
injection again at twelve distinct Doppler shift frequencies
between 200 Hz and 14 KHz.
A single switch on the phantom front panel allows selection of either
the onboard audio signal generator or an external signal source (passed
through the audio phasing board), such as a cassette deck, for pre-recorded
waveform injection. The complete phantom is pictured in figure 1.3.
Three parameters are of interest in characterisation of the phantom's performance
Tests were performed using a Phillips PM5133 audio function generator to produce the required audio frequencies and a Wavetek function generator emulated scanner signals with a radio frequency sine wave. As waveform injection is a priority for this phantom, all audio frequencies were piped through the Dynamic Imaging audio phasing board for this experiment. The onboard audio generator typically provides a constant 30 db sideband suppression and is primarily intended for frequency abscissa calibration checking.
Figure 1.4 shows the results of tests one and two above. Sideband
linearity is quantified by inputting an audio signal at typical operating
power levels into the phantom and measuring the injected sideband power level
relative to the noise floor. All power measurements were taken with the aid
of a Marconi Instruments model TF 2370 graphic spectrum analyser.
As supplied, the operational range of the audio phasing board is given as between 300 Hz and 7 kHz. It will be observed that sideband suppression is optimum, however, between 750 Hz and 6 kHz over which an excellent constant suppression of 52±2 dB is achieved. The error figure arises from the resolution of the spectral analyser display screen. Over this frequency range the lower sideband power steadily increased from 61 dB to 64 dB relative to the noise floor of the spectral analyser display, illustrating a remarkably high degree of linearity over the working range of the phantom.
In figure 1.5, the sideband suppression as measured using a typical
audio frequency of 4 kHz is plotted for eight several scanner
frequencies spanning the ESI phantom's operational range,
characterisation test 3 above.
Again a very high suppression of approximately 53 dB is obtained
over the operational range of the phantom. The slightly lower
suppression observed at 2 MHz is due to a fractionally higher inductance used
in the parallel tuned phasing circuit causing imperfect quadrature
phase shifting. This has since been rectified by changing to a lower
inductance coil for this filter.
Overall the phantom exceeds all previously published sideband
suppression figures. The built-in audio oscillator allows 100%
reproducable audio test signals to be injected for frequency abscissa
calibration checking (Figure 1.6) whilst any waveform lying within the
wide audio bandwidth of the phantom may be injected via the Dynamic Imaging
audio phasing board.
In order that possible variations in waveform indices between different scanners (and waveform index calculation packages) be detectable, the unavoidable experimental error induced by making the measurement must first be quantified. The nature of this error is identified by considering the experimental procedure. A pre-recorded waveform is first injected for a set number of cycles (time period) into the scanner under test and several cycles are chosen for the waveform index calculation package to operate on, as per clinical practice. A light pen attached to the Doptek scanner is used to select the start and finish of the chosen time period over which the waveform index calculation is to be made. The effect of small differences in the start and finish times on the resulting index is quantified by calculating the mean and standard deviation of the resulting waveform index upon repetition of this procedure several times.
Other intrinsic sources of error to the waveform index calculation must be minimised in order to make useful observations of operator induced error. The waveform generated by a computer controlled phantom is characterised by a highly reproduceable frequency envelope, and is thus suitable for the minimisation of errors associated with a variable waveform envelope between cycles. The reader should also recall that statistical variations in the power spectral density within an average cycle also play a part in determining the resulting waveform index error figure, but these are hopefully smaller than those induced by operator error. As a control, the experiment was also conducted on a simple, artificially generated waveform which has a constant spectral power density at identical phases of each cycle. (The control waveform was produced by frequency modulating a 4.65 kHz audio signal with another of approximately 1.5Hz so that the resulting frequency envelope sinusoidally oscillates between 2.5 kHz and 6.8 kHz, lying within the linear operational audio bandwidth of the electronic phantom.)
Finally, it has been experimentally observed that waveform index calculation packages are influenced by scanner gain settings. For completeness the above experimental procedure was repeated for three different scanner gain settings. These gain settings were similar to those routinely used in clinical practice with this scanner, especially the `medium' gain (table 1.1), with the highest gain used still producing a noise free spectrum.
For each of the three gain settings used (low, medium and high as defined in table 1.1) the mean pulsatility index (PI) was determined over a five cycle period using the Doptek's [ 15/16] percentile method. This was repeated six times to gain the mean and standard deviation figures shown in table 1.1 for the real waveform, and in table 1.2 for the control waveform depicted in figure 1.8.
| Doptek Gain | Spectral Description | P.I. | s PI |
|
| Low | Weak, completely unsaturated | - 1.94 | 0.03 | |
| Medium | Slightly saturated | -1.35 | 0.05 | |
| High | Highly saturated | - 1.33 | 0.04 | |
| Doptek Gain | Spectral Description | P.I. | sPI |
|
| Low | Weak, completely unsaturated | -0.823 | 0.004 | |
| Medium | On point of saturation | -0.825 | 0.005 | |
| High | Completely Saturated | -0.820 | 0.000 | |
A portable, all electronic Doppler phantom has been described. Use of a commercially obtained wide-band audio phase shifting board allows the injection of realistic waveforms, a facility allowing novel QA techniques such as the accuracy of waveform index calculation packages to be assessed. The design of this phantom results in exceptionally high sideband supression and a highly linear power spectral density injection characteristic at a discrete number of scanner operating frequencies which lie in the range of 2-10 MHz. The Edinburgh ESI phantom can also inject any of twelve internally generated, discrete simulated Doppler shift audio frequencies for fast testing of the Doppler frequency mapping operation of ultrasonic scanners. Such tests are free from large, angle-induced measurement errors as the Doppler shift is created electronically.
The phantom has been used to demonstrate that waveform index calculation package checking is quite feasable and can obtain highly reproduceable results as demonstrated in this paper using the PI package on a Doptek CW scanner. Although the realistic waveform was produced by a flow phantom, such phantoms have difficulty in maintaining waveform integrity (average power spectral density, envelope etc.) over time and are almost completely unportable. The ESI phantom therefore also provides an essential link enabling the same `captured' flow phantom waveform to be used again and again on many different scanners with the typical ease of use (portability, speed of setup) of an electronic device.